OPTILITE® FREELITE® KAPPA FREE KIT (X10)
Received Jul 6, 2026 · Event occurred Dec 17, 2025
Report 9614373-2026-00005 · MDR key 25718417
Device
Generic name
Optilite® Freelite® Kappa Free Kit (x10)
Manufacturer
The Binding Site LimitedModel number
LK016.10.OPTCatalog number
LK016.10.OPT
Lot number
530575
Product problems
- High Test Results
Patient
88 YR · Male
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
Narrative
Description of Event or Problem
THIS EVENT DID NOT OCCUR IN THE USA. THIS EVENT OCCURRED IN FINLAND ON A DEVICE (OPTILITE® FREELITE® KAPPA FREE KIT (X10), MODEL #: LK016.1 0.OPT) THAT IS SIMILAR TO THE DEVICES (OPTILITE FREELITE KAPPA FREE KIT MODEL#: LK016.OPT.A (510(K) NUMBER: K231290), OPTILITE FREELITE KAPPA FREE KIT (X10) MODEL#: LK016.10.OPT.A (510(K) NUMBER: K231290) THAT ARE MARKETED IN THE USA. ON 9 JUNE 2026, THE BINDING SITE (TBS) WERE MADE AWARE BY A CUSTOMER IN FINLAND OF AN ALLEGED FALSELY ELEVATED KAPPA FREELITE FREE LIGHT CHAIN (FLC) RESULT OF 1269 MG/L OBTAINED ON THE OPTILITE ANALYSER, RELEASED FROM THE LABORATORY ON (B)(6) 2025 FOR A PATIENT BEING INVESTIGATED FOR POSSIBLE MYELOMA. THIS RESULT WAS REPORTED TO THE CLINICIAN. DUE TO THE FALSELY ELEVATED RESULT, THE PATIENT UNDERWENT A BONE MARROW BIOPSY ON (B)(6) 2026. FOLLOWING FURTHER TESTING, SIGNIFICANTLY LOWER KAPPA FLC RESULTS WERE OBTAINED, INDICATING THAT THE INITIAL RESULT WAS ERRONEOUS. THERE HAVE BEEN NO REPORTS OF DEATH OR INDICATION OF COMPLICATIONS FOLLOWING THE BONE MARROW BIOPSY. TBS CONSIDERS AN UNNECESSARY BONE MARROW BIOPSY TO BE AN INVASIVE PROCEDURE AND CLASS IT AS A SERIOUS MEDICAL EVENT.
Additional Manufacturer Narrative
TO DATE, NO ASSAY PERFORMANCE ISSUES HAVE BEEN IDENTIFIED.