inforMED
InjuryOOE

ALLY ADAPTIVE CATARACT TREATMENT SYSTEM

Received Jul 6, 2026 · Event occurred Jun 17, 2026

Report 3009026057-2026-00030 · MDR key 25716595

Device

Generic name

Ally Adaptive Cataract Treatment System

Manufacturer

Lensar, Inc.

Model number

N/A

Catalog number

70-00050-001

Lot number

N/A

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • Corneal Perforation

Narrative

Additional Manufacturer Narrative

REVIEW OF PROCEDURE#: (B)(6) REVEALS EYE MOVEMENT ON FRAME 27 JUST AS THE ARCUATE INCISION BEGINS AND ARCUATE COMPLETION ON FRAME 34. NO INDICATION OF FULL THICKNESS INCISION IS SEEN. RECOMMEND REVIEW OF ARCUATE INCISION OPENING TECHNIQUE. SYSTEM FUNCTIONED AS DESIGNED. NO FURTHER CLINICAL FOLLOW UP REQUIRED.

Description of Event or Problem

ON (B)(6) 2026, DOCTOR (B)(6) REPORTED TO CAS THAT THEY EXPERIENCED A FULL THICKNESS ARCUATE INCISION DURING PROCEDURE ID#: (B)(6).