inforMED
MalfunctionFZP

LIGACLIP

Received Jul 6, 2026 · Event occurred May 20, 2026

Report 3005075853-2026-05247 · MDR key 25716052

Device

Generic name

Clip, Implantable

Catalog number

ER320

Lot number

501D90

Product problems

  • Failure to Fire

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

Additional Manufacturer Narrative

(B)(4). DATE SENT: 7/6/2026. D4: BATCH # Z70092. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO EES FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE ER320 DEVICE WAS RECEIVED FOR EVALUATION, AND UPON INSPECTION, THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION. IN ADDITION, THE OPENED PACKAGING WAS RETURNED ALONG WITH THE INSTRUMENT. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS SUBJECTED TO FUNCTIONAL EVALUATION. DURING FIRING, THE REMAINING FIFTEEN (15) CLIPS WERE EXPELLED DUE TO THE COMPROMISED CONDITION OF THE JAWS. SUBSEQUENTLY, THE INSTRUMENT REACHED LOCKOUT AS INTENDED. THE EVENT REPORTED WAS CONFIRMED AND IT IS RELATED TO IMPROPER USE OF THE DEVICE. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE YIELDED JAWS MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP, PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION, OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. AS PART OF EES QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. DEVICE HISTORY REVIEW: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH Z70092 NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.