AMISTEM-P COLLARED
Received Jul 6, 2026 · Event occurred Jun 10, 2026
Report 3005180920-2026-00622 · MDR key 25714736
Device
Generic name
Amistem-p Collared Lat. Size0
Manufacturer
Medacta International SaModel number
01.18.440Lot number
2520365
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Unknown
- Bone Fracture(s)
Narrative
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 11 JUNE 2026 STEM: AMISTEM-P COLLARED 01.18.440 AMISTEM-P COLLARED LAT. SIZE 0 (K173794) LOT 2520365: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26-FEB-2026. EXPIRATION DATE: 08-FEB-2031. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE, NO DEFINITIVE ROOT CAUSE CAN BE ESTABLISHED. THE ABSENCE OF X-RAYS AND FURTHER CLINICAL DETAILS DOES NOT ALLOW A MORE PRECISE CONCLUSION. THE DEVICE HISTORY RECORD REVIEW DID NOT INDICATE ANY POTENTIALLY RELATED MANUFACTURING ISSUE, AND THERE IS NO INDICATION THAT A DEVICE-RELATED ISSUE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem
AT ABOUT 1 WEEK FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A FRACTURED FEMUR AND THE CAUSE IS UNKNOWN. THERE WERE NO INDICATIONS OF TRAUMA.THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH COMPETITOR DEVICES AND REPLACED THE DOUBLE MOBILITY LINER WITH A LINER OF THE SAME SIZE. THE SURGERY WAS COMPLETED SUCCESSFULLY.