UTK
Received Jul 5, 2026 · Event occurred Nov 11, 2025
Report 3011164092-2025-00059 · MDR key 25712524
Device
Generic name
Infrared Heating Pad
Manufacturer
Utk Technology Shen Zhen LimitedModel number
H11S2Catalog number
H11S2
Lot number
UNKNOWN
Product problems
- Overheating of Device
Patient
39 YR · Unknown
- Burn(s)
Narrative
Additional Manufacturer Narrative
THIS ADVERSE EVENT INFORMATION WAS SOURCED FROM A PUBLIC AMAZON CUSTOMER REVIEW (REVIEW ID: (B)(4) DATED ON (B)(6) 2025. MODEL: H11S2, SMOKED DURING NORMAL REPEATED USE. THE PLUG AREA TEMPERATURE EXCEEDED 250°F MEASURED BY HEAT GUN, WHICH CAUSED SKIN BURN INJURY TO THE END USER AND SCORCHED BED SHEETS. RECURRENCE OF THIS OVERHEATING FAILURE CREATES POTENTIAL FIRE HAZARD. THE MANUFACTURER CANNOT CONTACT THE ANONYMOUS CUSTOMER TO RETRIEVE THE FAILED UNIT OR COLLECT FURTHER CLINICAL AND DEVICE INSPECTION DETAILS. NO CORRECTIVE OR REMEDIAL ACTIONS HAVE BEEN INITIATED TO DATE. THE PRODUCT IS 510(K) EXEMPT REUSABLE THERMAL THERAPY DEVICE. SINCE SPECIFIC HEATING ELEMENT COMPONENT CODE IS UNAVAILABLE IN THIS SOFTWARE LIBRARY, HEAT EXCHANGER IS SELECTED AS THE EQUIVALENT THERMAL COMPONENT CATEGORY TO REPRESENT THE INTERNAL HEATING WIRE OF THE THERMAL THERAPY PAD WHICH CAUSED OVERHEATING FAILURE.
Description of Event or Problem
MODEL: H11S2, INCIDENT DATE: ON (B)(6) 2025. DEVICE SMOKED DURING USE, PLUG AREA TEMPERATURE OVER 250°F MEASURED BY HEAT GUN, CAUSED USER SKIN BURN AND BURNT BED SHEETS. FIRE HAZARD EXISTS IF FAILURE RECURS. THIS FEEDBACK WAS OBTAINED FROM PUBLIC AMAZON CUSTOMER REVIEW (REVIEW ID: (B)(4), AND THE CONSUMER CANNOT BE CONTACTED FOR FURTHER FOLLOW-UP INFORMATION. INTERNAL CASE NO. (B)(4).