BREATHING CIRCUIT SET
Received Jul 3, 2026 · Event occurred Jun 16, 2026
Report 3001421318-2026-00405 · MDR key 25704627
Device
Generic name
Breathing Circuit Set
Manufacturer
Hamilton Medical AgModel number
260128Catalog number
260128
Lot number
202500
Product problems
- Obstruction of Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
HAMILTON MEDICAL AG REFERENCE NUMBER: (B)(4). INVESTIGATION IS ONGOING. HMAG REFERENCE NUMBER: (B)(4) FDA REFERENCE: (B)(4).
Description of Event or Problem
IT WAS REPORTED TO HAMILTON MEDICAL AG, THAT: "REPORTING A KNOWN ISSUE WITH THE MANUFACTURING OF THE BREATHING CIRCUIT FOR HAMILTON T1. AFTER MOVING PATIENT FROM BEDSIDE VENTILATOR TO THE PORTABLE ONE, THE HAMILTON T1 STARTED ALARMING AND WAS UNABLE TO VENTILATE THE PATIENT. THE RESIDENT ANAESTHETIST CHANGED THE EXPIRATORY VALVE OF THE CIRCUIT WITH ANOTHER ONE AND THE ISSUE WAS RESOLVED. ACTIONS TAKEN AS PER THE FSN INSTRUCTIONS BY MANUFACTURER." PATIENT INVOLVEMENT AND MEDICAL INTERVENTION, BUT NO PATIENT, USER OR THIRD PARTY HARM WAS REPORTED.
Remedial action
- Notification