inforMED
InjuryQHR

Anti-Human Globulin Anti-IgG (Rabbit) (Green) IH-Card AHG Anti-IgG

Received Jul 3, 2026 · Event occurred May 28, 2026

Report 9610824-2026-00018 · MDR key 25701449

Device

Generic name

Ih-card Ahg Anti-igg (rabbit); 288 Cards

Catalog number

813422100

Lot number

9511030

Product problems

  • False Negative Result

Patient

Unknown

  • Hemolysis
  • Hypersensitivity/Allergic reaction
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

Narrative

Description of Event or Problem

CUSTOMER REPORTED THAT A PATIENT EXPERIENCED A TRANSFUSION REACTION. THE ANTIBODY SCREEN IN GEL USING IH-CARD AHG ANTI-IGG (PN 813422100, LOT 9511030, EXPIRY 7/12/2026) FOR THE PRE-TRANSFUSION SAMPLE WAS NEGATIVE. THE ANTIBODY SCREEN IN GEL USING IH-CARD AHG ANTI-IGG (PN 813422100, LOT 9511030, EXPIRY 7/12/2026) FOR THE POST-TRANSFUSION SAMPLE WAS NEGATIVE. CUSTOMER REPORTED SHE RAN THE PRE-TRANSFUSION AND POST-TRANSFUSION SAMPLES ON THE IMMUCOR ECHO AND IDENTIFIED AN ANTI-JKA.

Additional Manufacturer Narrative

INITLAL REPORT.