inforMED
InjuryJCW

AMS 700 CXR

Received Jul 2, 2026 · Event occurred Jun 4, 2026

Report 2124215-2026-35638 · MDR key 25699787

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404012

Catalog number

72404012

Lot number

1100077173

Product problems

  • Mechanical Problem
  • Material Puncture/Hole

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. BASED ON THE INFORMATION AVAILABLE THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED. THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION. MODEL NUMBER/CATALOG NUMBER: 72404310 SERIAL NUMBER: N/A BATCH/LOT NUMBER: 1100080995 MODEL/CATALOG DESCRIPTION: PUMP MS UNIQUE IDENTIFIER (UDI) #: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT FUNCTIONING PROPERLY. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH A CRACK WAS FOUND IN THE RIGHT CYLINDER CONNECTING TUBE. THE RIGHT CYLINDER AND THE PUMP WERE REMOVED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.