Generic name

Catalog Kit (primary Device Kit)

Catalog number

689852/JPB

Lot number

C3030703

Product problems

  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SUSPECTED DEVICE HAS BEEN RETURNED FOR EVALUATION. THE COMPLAINT IS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE ATTRIBUTED TO SURFACE CONTAMINATION ON THE LUER LOCK OF THE SYRINGE, RESULTING TO POOR ADHESION OF THE 2 PARTS. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE PRODUCT HISTORY REVIEW WAS ANALYZED, AND NO EXCEPTION DOCUMENTS WERE FOUND. NONCONFORMANCES OF THIS NATURE ARE MONITORED BY THE OPERATION AND ENGINEERING TEAM. THIS COMPLAINT WILL BE USED TO TREND FOR SIMILAR COMPLAINTS.

Description of Event or Problem

CUSTOMER REPORTED SUSPECTED LEAK (NOT A BLOOD LEAK). NO PATIENT INJURY.