MalfunctionOUG
AQURE
Received Jul 2, 2026 · Event occurred Jun 19, 2026
Report 3002807968-2026-00068 · MDR key 25689077
Device
Product problems
- Communication or Transmission Problem
- Data Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, THE CUSTOMER EXPERIENCED SOME OF THE ABL90 ANALYZERS COULD NOT SEND THE DATA TO AQURE (SERIAL NUMBER: (B)(6), WITH THE FOLLOWING ERROR MESSAGE IN THE LOG FILES: "EXCEPTION IN REPORTPATIENTOBSERVATION MESSAGE AUTOMAPPER.AUOMMAPPERMAPPINGEXCEPTION: ERROR MAPPING TYPES.
Additional Manufacturer Narrative
DATE OF EVENT IS ESTIMATED BASED ON MANUFACTURER AWARENESS DATE. SERIAL NUMBER OF DEVICE: (B)(6).