ELECSYS ANTI-TPO
Received Jul 1, 2026 · Event occurred Jun 10, 2026
Report 1823260-2026-02664 · MDR key 25681630
Device
Generic name
System, Test, Thyroid Autoantibody
Manufacturer
Roche DiagnosticsCatalog number
07026935190
Lot number
92523001
Product problems
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COBAS E 801 MSB/MSBL SERIAL NUMBER IS (B)(6). THE PATIENT'S SAMPLE HAS BEEN RECEIVED. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS ANTI-TPO RESULTS FOR 1 PATIENT ON A COBAS E 801 MSB/MSBL MODULE. THE INITIAL RESULT WAS 125 IU/ML (REFERENCE INTERVAL: <34IU/ML), WHICH IS CLINICALLY INTERPRETED AS POSITIVE. THE RESULT ON ANOTHER COBAS E 801 WAS 157 IU/ML, CONFIRMING THE ELEVATED LEVEL AS POSITIVE. THE RESULT USING A SNIBE MAGLUMI X8 ANALYZER WAS 0.061 IU/ML (REFERENCE INTERVAL: <10 IU/ML, WHICH IS CLINICALLY INTERPRETED AS NEGATIVE. THE RESULT USING A SIEMENS ATELLICA IM ANALYZER WAS 28.9 IU/ML (REFERENCE INTERVAL: <60 IU/ML), WHICH IS CLINICALLY INTERPRETED AS NEGATIVE. THE RESULT USING A BECKMAN COULTER UNICEL DXI 800 ANALYZER WAS 1.19 IU/ML (REFERENCE INTERVAL: <9 IU/ML), WHICH IS CLINICALLY INTERPRETED AS NEGATIVE. THE RESULT USING AN ABBOTT ALINITY I / I2000SR ANALZYER WAS 0.33 IU/ML (REFERENCE INTERVAL: <5.61 IU/ML), WHICH IS CLINICALLY INTERPRETED AS NEGATIVE. THE CUSTOMER IS QUESTIONING THE INITIAL RESULTS BASED ON THE CLINICAL PICTURE, NEGATIVE IMAGING, AND UNANIMOUS NEGATIVE RESULTS FROM FOUR OTHER MAJOR MANUFACTURERS. THE PATIENT'S THYROID ULTRASOUND SHOWED NO ABNORMALITIES.