AMS 700 CX MS PUMP
Received Jul 1, 2026 · Event occurred Feb 1, 2026
Report 2124215-2026-35218 · MDR key 25681201
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404231Catalog number
72404231
Product problems
- Fluid/Blood Leak
- Inflation Problem
- Material Puncture/Hole
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DETAILED PRODUCT INFORMATION WAS NOT PROVIDED TO BSC. WE COMPLETED A GOOD FAITH EFFORT TO OBTAIN THE INFORMATION. BECAUSE THE PRODUCT IS UNKNOWN AT THIS TIME, WE ARE UNABLE TO PROVIDE THE COMPLETE UNIQUE IDENTIFIER (UDI) # AND OTHER PRODUCT SPECIFIC INFORMATION. IF ADDITIONAL DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED A SENSATION OF INTERNAL MOISTURE WITHIN THE PENIS AND SCROTUM AND WAS UNABLE TO INFLATE THE DEVICE. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH DAMAGE AND DELAMINATION OF THE LEFT CYLINDER WERE IDENTIFIED IN ITS DISTAL HALF. THE CYLINDERS AND PUMP WERE REMOVED, AND FOLLOWING DISCONNECTION OF THE RESERVOIR, ASPIRATION CONFIRMED THAT NO FLUID WAS PRESENT WITHIN THE SYSTEM. SUBSEQUENTLY, NEW CYLINDERS AND A NEW PUMP WERE IMPLANTED AND CONNECTED TO THE ORIGINAL RESERVOIR, WHICH WAS LEFT IN SITU. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.