inforMED
MalfunctionCFR

GLUCOSE HK

Received Jul 1, 2026 · Event occurred Jun 11, 2026

Report 1823260-2026-02639 · MDR key 25677000

Device

Generic name

Hexokinase, Glucose

Manufacturer

Roche Diagnostics

Catalog number

04657527190

Lot number

94209701

Product problems

  • Non Reproducible Results

Patient

36 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE C 111 ANALYZER IS (B)(4). CALIBRATION AND CONTROLS WERE ACCEPTABLE. A GENERAL REAGENT ISSUE CAN BE RULED OUT. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH GLUCOSE HK ON A COBAS C 111 ANALYZER. THE SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 208.84 MG/DL AND IT REPEATED AS 107.23 MG/DL.