inforMED
MalfunctionDAD

TINA-QUANT HAPTOGLOBIN VER.2

Received Jul 1, 2026 · Event occurred Jun 5, 2026

Report 1823260-2026-02622 · MDR key 25676493

Device

Generic name

Haptoglobin, Antigen, Antiserum, Control

Manufacturer

Roche Diagnostics

Catalog number

08106045190

Lot number

88577601

Product problems

  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE C503 ANALYZER SERIAL NUMBER WAS (B)(6).

Description of Event or Problem

THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR TINA-QUANT HAPTOGLOBIN VER.2 (HAPT2) ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS > 5.80 G/L WITH A DATA FLAG. THE REPEAT RESULT WAS >5.82 G/L WITH A DATA FLAG. THE REPEATED RESULT WITH A 1:3 DILUTION WAS 0.0388 G/L. THE SAMPLE WAS REPEATED, AND THE RESULT WAS 1.83 G/L WITH A DATA FLAG. A 1:3 MANUAL DILUTION WAS PERFORMED WITH A CALCULATED RESULT OF 5.49 G/L. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER SUSPECTS A SAMPLE-SPECIFIC INTERFERENCE.