inforMED
MalfunctionLLZ

INTELLISPACE CARDIOVASCULAR

Received Jul 1, 2026 · Event occurred Apr 30, 2025

Report 3003768277-2026-101212 · MDR key 25675210

Device

Generic name

Cardiovascular Image And Information Management System

Catalog number

830089

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

PHILIPS IDENTIFIED A SOFTWARE DEFECT INTERNALLY IN ISCV (INTELLISPACE CARDIOVASCULAR) WHEREIN THERE IS NOT A WAY TO HIGHLIGHT STAT PATIENTS. THE ISSUE WAS IDENTIFIED INTERNALLY AND WAS NOT IN CLINICAL USE AT THE TIME OF EVENT. BASED ON THE NEW INFORMATION THAT PHILIPS BECAME AWARE, THIS HAS BEEN DETERMINED TO BE A REPORTABLE MALFUNCTION. UNDER SPECIFIC CIRCUMSTANCES, IT COULD POTENTIALLY LEAD TO DELAYED DIAGNOSIS. GIVEN THE DEFECT¿S NATURE AND ITS POTENTIAL IMPACT ON CLINICAL DECISION-MAKING IF IT WERE TO RECUR, IT IS DETERMINED THAT THIS CONSTITUTES A REPORTABLE MALFUNCTION.