inforMED
InjuryJCW

AMS 700 LGX MS Pump

Received Jun 30, 2026 · Event occurred Jun 4, 2026

Report 2124215-2026-34864 · MDR key 25673425

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404251

Catalog number

72404251

Lot number

1100626351

Product problems

  • Inadequacy of Device Shape and/or Size
  • Malposition of Device

Patient

74 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED WITH THE PUMP RIDING HIGH. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN CONFIRMED THE PUMP WAS POSITIONED CLOSE TO THE SHAFT WITH INSUFFICIENT TUBING. THE DEVICE WAS EXPLANTED AND REPLACED, AND A NEW PUMP WAS POSITIONED APPROPRIATELY WITHIN THE SCROTUM. THERE WERE NO PATIENT COMPLICATIONS.

Additional Manufacturer Narrative

MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100786176. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)(4).