MERGE PACS
Received Jun 30, 2026 · Event occurred May 22, 2026
Report 2183926-2026-00003 · MDR key 25668272
Device
Generic name
Medical Image Management And Processing System
Manufacturer
Merge HealthcareModel number
9.4 P02Product problems
- Computer Software Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN INVESTIGATION HAS BEEN COMPLETED AND THE ROOT CAUSE IS DUE TO A SOFTWARE DEFECT INTRODUCED IN MERGE PACS 9.4. ALTHOUGH THERE ARE SEVERAL VISUAL CUES TO ALERT USERS TO THE PRESENCE OF NEW IMAGES WHEN VIEWING A STUDY, UNDER HIGH-VOLUME CLINICAL WORKLOADS THERE IS A POTENTIAL RISK THAT RADIOLOGISTS MAY READ AND INTERPRET IMAGES FOR THE INCORRECT PATIENT IN THE CURRENT STUDY CONTEXT. AN IMMEDIATE WORKAROUND IS BEING DEPLOYED TO AFFECTED CUSTOMERS TO DISABLE THE FEATURE THROUGH A CONFIGURATION CHANGE, AND A SOFTWARE UPDATE IS UNDER DEVELOPMENT TO ELIMINATE THE ROOT CAUSE IN ALL AFFECTED VERSIONS.
Description of Event or Problem
ON (B)(6) 2026, MERGE HEALTHCARE TECHNICAL SUPPORT RECEIVED A COMPLAINT FROM A SITE'S RADIOLOGY IT/PACS ADMINISTRATOR REPORTING THAT, WHEN LAUNCHING A STUDY IN MERGE PACS, IMAGES FROM TWO DIFFERENT PATIENTS WERE DISPLAYED TOGETHER. THE COMPLAINT WAS INITIALLY INVESTIGATED AND THE REPORTED BEHAVIOR COULD NOT BE REPRODUCED. ON (B)(6) 2026, A SECOND COMPLAINT DESCRIBING SIMILAR BEHAVIOR WAS RECEIVED. SUBSEQUENT INVESTIGATION CONFIRMED THAT, UNDER CERTAIN CONDITIONS, THE SOFTWARE COULD DISPLAY IMAGES ASSOCIATED WITH DIFFERENT PATIENTS WITHIN THE SAME VIEWING SESSION. THE ISSUE WAS DETERMINED TO BE A SOFTWARE MALFUNCTION. NO PATIENT INJURY OR ADVERSE CLINICAL OUTCOME WAS REPORTED.
Remedial action
- Recall