MERGE PACS
Received Jun 30, 2026 · Event occurred Jun 4, 2026
Report 2183926-2026-00001 · MDR key 25668246
Device
Generic name
Medical Image Management And Processing System
Manufacturer
Merge HealthcareModel number
9.4 P02Product problems
- Computer Software Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN INVESTIGATION HAS BEEN COMPLETED AND THE ROOT CAUSE IS DUE TO A SOFTWARE DEFECT INTRODUCED IN MERGE PACS 9.4. ALTHOUGH THERE ARE SEVERAL VISUAL CUES TO ALERT USERS TO THE PRESENCE OF NEW IMAGES WHEN VIEWING A STUDY, UNDER HIGH-VOLUME CLINICAL WORKLOADS THERE IS A POTENTIAL RISK THAT RADIOLOGISTS MAY READ AND INTERPRET IMAGES FOR THE INCORRECT PATIENT IN THE CURRENT STUDY CONTEXT. AN IMMEDIATE WORKAROUND IS BEING DEPLOYED TO AFFECTED CUSTOMERS TO DISABLE THE FEATURE THROUGH A CONFIGURATION CHANGE, AND A SOFTWARE UPDATE IS UNDER DEVELOPMENT TO ELIMINATE THE ROOT CAUSE IN ALL AFFECTED VERSIONS.
Description of Event or Problem
ON (B)(6) 2026, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A CUSTOMER. THE SITE'S PACS ADMINISTRATOR REPORTED THAT THUMBNAIL IMAGES FROM A PREVIOUS PATIENT'S STROKE PROTOCOL WERE DISPLAYED IN THE THUMBNAIL BAR FOR A SUBSEQUENT PATIENT. NO INJURY WAS REPORTED. IF THE ISSUE WERE TO RECUR, IT COULD POTENTIALLY LEAD TO INCORRECT CLINICAL INTERPRETATION.
Remedial action
- Recall