GIRAFFE OMNIBED CARESTATION
Received Jun 30, 2026 · Event occurred Jun 5, 2026
Report 3008729547-2026-00052 · MDR key 25668093
Device
Generic name
Incubator, Neonatal
Manufacturer
Critikon De Mexico S. De R.l. De C.v.Model number
CS1Catalog number
NA
Lot number
NA
Product problems
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR THIS HEATER DOOR ISSUE PER 21 CFR 806 ON (B)(6) 2024. THE FDA RECALL NUMBER IS Z-0047-2025 & Z-0048-2025. CUSTOMERS WERE SENT A LETTER EXPLAINING THE ISSUE AND INSTRUCTIONS FOR INSPECTING, INSTALLING, AND TIGHTENING THE SCREW THAT SECURES THE HEATER DOOR. GEHC WILL REPLACE ALL AFFECTED UNITS. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.
Description of Event or Problem
THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH THE HEATER DOOR THAT INDICATES IT MAY BE IMPACTED BY THE ISSUE BEING ADDRESSED AS PART OF CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON (B)(6) 2024 (Z-0047-2025 & Z-0048-2025). THERE WAS NO PATIENT INVOLVEMENT.
Remedial action
- Recall