3200 HYPERBARIC CHAMBER

Received May 4, 2012 · Event occurred Apr 3, 2012

Report 2020676-2012-00017 · MDR key 2566545

Device

Generic name

Chamber, Hyperbaric

Model number

3200

Catalog number

3200

Product problems

  • Output above Specifications
  • Increase in Pressure

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

TECHNICIAN IS SCHEDULED FOR SITE VISIT TO EVALUATION AND PROVIDE SERVICE. REPORTED COMPLAINT WAS FOUND DURING REGULARLY SCHEDULED TESTING. NO PATIENT INVOLVEMENT WHEN COMPLAINT WAS FOUND. A FOLLOW-UP SUBMISSION REPORT OF FINDINGS WILL BE PROVIDED ONCE THE EVALUATION OF THE DEVICE IS CONDUCTED.

Description of Event or Problem

THE CHAMBER IS COMPRESSING FASTER THAN SPECIFIED. AT A RATE OF 1 THE CHAMBER IS PRESSURIZING A 41 SEC/PSI. WHEN THE PRESSURE RATE IS SET TO 3 THE COMPRESSION IS AT THE HIGH END OF THE SPECIFICATION OF 50SEC/PSI. THE TECHNICIAN, STATED THAT HE IS KEEPING AN EYE ON THE PRESSURE AND TAKING THE NECESSARY STEPS/PRECAUTIONS FOR PATIENT SAFETY.