inforMED
MalfunctionJGJ

ARCHITECT Magnesium

Received Jun 30, 2026 · Event occurred May 15, 2026

Report 3005094123-2026-00321 · MDR key 25665322

Device

Generic name

Photometric Method, Magnesium

Catalog number

03P68-24

Lot number

83306UD00

Product problems

  • High Test Results

Patient

29 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem

THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED BY THE ARCHITECT C4000 PROCESSING MODULE FOR A PATIENT. THE RESULTS WERE NOT RELEASED. THE SAMPLE WAS REPEATED, AND A NORMAL RESULT WAS OBTAINED. THE FOLLOWING DATA WERE PROVIDED: NORMAL RANGE: 1.6 TO 2.6 MG/DL (B)(6) 2026 SID (B)(6). INITIAL RESULT = 9.14 MG/DL REPEAT RESULT ON THE SAME INSTRUMENT = 3.96 MG/DL REPEAT RESULT ON ANOTHER INSTRUMENT (B)(6) = 1.9 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.