AMS 700 LGX? TENACIO? PUMP
Received Jun 29, 2026 · Event occurred Nov 8, 2024
Report 2124215-2026-34621 · MDR key 25661316
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404451Catalog number
72404451
Lot number
1100511403
Product problems
- Unintended Deflation
Patient
62 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100589277. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #:(B)(4).
Description of Event or Problem
IT WAS REPORTED THAT FOLLOWING INITIAL ACTIVATION OF THIS INFLATABLE PENILE PROSTHESIS (IPP); THE DEVICE EXHIBITED SPONTANEOUS DEFLATION DURING INTERCOURSE. THE PATIENT CONSULTED THEIR PHYSICIAN AND WAS GUIDED TO WAIT FOR AN ADDITIONAL PERIOD BEFORE ATTEMPTING USE OF THE DEVICE. APPROXIMATELY 3-4 MONTHS AFTER IMPLANT, THE PATIENT WAS GIVEN CLEARANCE TO USE THE DEVICE ONCE MORE AND THE PATIENT EXPERIENCED SPONTANEOUS DEFLATION AGAIN. A SURGICAL PROCEDURE WAS THEN PERFORMED DURING WHICH THE IPP WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.