AMS 700 CXR MS PUMP

Received Jun 29, 2026 · Event occurred Feb 1, 2026

Report 2124215-2026-34615 · MDR key 25661237

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404262

Catalog number

72404262

Lot number

1100375515

Product problems

  • Inflation Problem
  • Mechanical Problem

Patient

38 YR · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. BASED ON THE INFORMATION AVAILABLE THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED. THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED INFLATION DIFFICULTIES DURING PUMPING. A SURGICAL INTERVENTION WAS PERFORMED, DURING WHICH A SURROGATE RESERVOIR TEST WAS CONDUCTED. BACK PRESSURE WAS OBSERVED IN THE SURROGATE SYRINGE DURING PUMPING, WITH NO FLUID TRANSFER THROUGH THE PUMP, INDICATING A SUSPECTED PUMP ISSUE. THE PUMP WAS RESET; HOWEVER, NO IMPROVEMENT WAS NOTED. THE RESERVOIR WAS INSPECTED AND CONFIRMED TO STILL CONTAIN FLUID. THE CYLINDERS AND REAR TIP EXTENDERS (RTES) WERE EXPLANTED. NEW CX CYLINDERS AND RTES WERE IMPLANTED AND CONNECTED TO THE EXISTING RESERVOIR. NO PATIENT COMPLICATIONS WERE REPORTED.