SURPASS EVOLVE 4.0MM X 17MM - CE

Received Jun 29, 2026 · Event occurred Jun 8, 2026

Report 3008881809-2026-00409 · MDR key 25660803

Device

Generic name

Intracranial Aneurysm Flow Diverter

Catalog number

FD40017

Lot number

26124673

Product problems

  • Difficult to Remove

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE SUBJECT STENT ENCOUNTERED RESISTANCE AT THE HUB OF THE FIRST MICROCATHETER. THE MICROCATHETER WAS REPLACED WITH THE NEW ONE, AND THE SUBJECT STENT WAS DELIVERED SMOOTHLY TO THE TARGET POSITION. THE SUBJECT STENT FAILED TO OPEN AT A VASCULAR CURVE. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE SUBJECT STENT BUT ENCOUNTERED SIGNIFICANT RESISTANCE. AFTER MULTIPLE ATTEMPTS, THE SUBJECT STENT COULD NOT BE RETRIEVED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

Additional Manufacturer Narrative

THIS IS 2ND OF 2 MDRS.