CUSTOM PROXIMAL HUMERUS

Received Jun 29, 2026 · Event occurred Feb 2, 2026

Report 3014833750-2026-00020 · MDR key 25654072

Device

Generic name

Custom Proximal Humerus

Manufacturer

Restor3d, Inc.

Model number

14726-110

Product problems

  • Detachment of Device or Device Component
  • Device Dislodged or Dislocated

Patient

NA · Unknown

  • Joint Dislocation

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION REVEALED A POSSIBLE DIMENSIONAL DISCREPANCY BETWEEN THE HUMERAL IMPLANT AND THE LINER IMPLANT THAT MAY HAVE IMPACTED THE FULL SEATING OF THE POLYETHYLENE LINER. THE LINER WAS ULTIMATELY CEMENTED INTO THE HUMERAL IMPLANT AND THERE IS LITTLE EXPECTED FURTHER RISK OF DISASSOCIATION BETWEEN THE COMPONENTS IN QUESTION.

Description of Event or Problem

A CUSTOM PROXIMAL HUMERUS RECONSTRUCTION IN CONJUNCTION WITH A REVERSE TOTAL SHOULDER HAD TO BE REVISED A SECOND TIME DUE TO POLYETHYLENE LINER DISASSOCIATION. THE INITIAL DISASSOCIATION EVENT INVOLVED THE PATIENT REPORTING MINOR TRAUMA, REACHING INTO GLOVE BOX AS CAR STOPPED WHICH CAUSED THE DISASSOCIATION. FOLLOWING INITIAL REVISION SURGERY, THE PATIENT DISLOCATED AGAIN PROMPTING THE SURGEON TO REVISE AGAIN.