inforMED
InjuryKWS

CUSTOM PROXIMAL HUMERUS

Received Jun 29, 2026 · Event occurred Feb 2, 2026

Report 3014833750-2026-00021 · MDR key 25654071

Device

Generic name

Custom Proximal Humerus

Manufacturer

Restor3d, Inc.

Model number

14580-110

Product problems

  • Detachment of Device or Device Component
  • Device Dislodged or Dislocated

Patient

NA · Unknown

  • Joint Dislocation

Narrative

Description of Event or Problem

A CUSTOM PROXIMAL HUMERUS RECONSTRUCTION IN CONJUNCTION WITH A REVERSE TOTAL SHOULDER REPLACEMENT HAD TO BE REVISED DUE TO POLYETHYLENE LINER DISASSOCIATION. THE PATIENT EXPERIENCED MINOR TRAUMA PLACING WEIGHT ONTO THE OPERATIVE ARM WHEN DISASSOCIATION OCCURRED.

Additional Manufacturer Narrative

THE INVESTIGATION HAD A DIMENSIONAL MEASUREMENT WHICH PROHIBITED FULL SEATING OF THE POLYETHYLENE LINER. CONTAINMENT HAS BEEN ADDRESSED AND NO OTHER DEVICES ARE IMPACTED.