inforMED
MalfunctionCAD

VAPOR 2000

Received Jun 29, 2026 · Event occurred May 25, 2026

Report 9611500-2026-00326 · MDR key 25650431

Device

Generic name

Anesthesia Unit Vaporizers

Model number

M35054

Product problems

  • Leak/Splash

Patient

NA · Unknown

  • Insufficient Information

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION HAS JUST STARTED, RESULTS WILL BE PROVIDED IN A FOLLOW-UP REPORT.

Description of Event or Problem

IT WAS REPORTED THAT THE ANESTHETIC AGENT SPRAYED OUT OF THE VAPORIZER DURING AN OPERATOR ATTEMPT TO FILL IT UP. AS PER REPORT, THE OPERATOR WAS NOT EXPOSED TO THE SUBSTANCE. REMARK: THE PARTICULAR AGENT INTENDED FOR USE WITH THIS DEVICE HAS A BOILING POINT OF 58°C / 136.4°F - IT SHOULD THUS NOT SPRAY OUT OF THE VAPORIZER AT ROOM TEMPERATURE. AT THE MOMENT, THE MANUFACTURER CANNOT EXCLUDE THE PRESENCE OF A MALFUNCTION OR USE ERROR AND THUS, THIS REPORT IS FILED IN ABUNDANCE OF CAUTION.