FLOWERS MANDIBULAR GLOVE
Received Jun 26, 2026 · Event occurred Jun 2, 2026
Report 2028924-2026-00003 · MDR key 25638912
Device
Product problems
- Break
Patient
39 YR · Unknown
- Failure of Implant
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ACTUAL DEVICE WAS RETURNED AND THE DEVICE SHOWED APPARENT PHYSICIAN MODIFICATION VIA CARVING/TRIMMING ALONG THE SUPERIOR EDGE AND LATERAL ARMS. IN ADDITION, THREE OTHER OPENINGS WERE FOUND WITH EDGE PATTERNS CHARACTERISTIC OF SHARP INSTRUMENT DAMAGE, THOUGH OPENINGS MAY HAVE BEEN EXTENDED VIA TEARING. PRODUCT LABELING WARNS THAT UTMOST CARE SHOULD BE TAKEN TO PREVENT IMPLANT DAMAGE DURING HANDLING OR SURGERY. LABELING ALSO INDICATES THAT THE PROPER SIZE AND SHAPE OF IMPLANTS MUST BE DETERMINED FOR THE INDIVIDUAL PATIENT BY THE SURGEON.
Description of Event or Problem
COMPLAINT REPORTED THAT THE DOCTOR COULD NOT GET THE DEVICE TO LIE FLUSH, AND WHILE HE WAS WORKING ON IT, IT SPLIT IN THE MIDDLE. ADDITIONAL INFORMATION WILL BE REQUESTED. THE DOCTOR PROCEEDED WITH ANOTHER DEVICE.