inforMED
MalfunctionHTJ

DEPTH GAUGE

Received Jun 26, 2026 · Event occurred Jan 1, 2026

Report 8030965-2026-06643 · MDR key 25636901

Device

Generic name

Depth Gauge

Manufacturer

Synthes Gmbh

Catalog number

319.006

Lot number

400389

Product problems

  • Detachment of Device or Device Component
  • Material Twisted/Bent

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: H3, H6: A PHOTO INVESTIGATION WAS COMPLETED: THE PHOTO INVESTIGATION REVEALED THAT THE TIP OF THE DEPTH GAUGE WAS OBSERVED BROKEN. THE TIP ITSELF IS NOT VISIBLE, THEREFORE, WEAR AND FELL APART CANNOT BE CONFIRMED. DUE TO THE BROKEN CONDITION, HOWEVER, THE COMPLAINT CONDITION CAN BE CONFIRMED. BASED ON THE AVAILABLE EVIDENCE, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Description of Event or Problem

IT WAS REPORTED THE DEPTH GAUGE HOOK HAD SOME WEAR AND TEAR NOTICED BEFORE THE PROCEDURE. WHEN TAKING THE SILVER THREAD ON PROTECTION SLEEVE OFF THE METAL HOOK OF THE DEPTH GAUGE FELL OFF.