TARGET XL 360 SOFT 6MM X 20CM
Received Jun 26, 2026 · Event occurred Jun 4, 2026
Report 3008881809-2026-00403 · MDR key 25636561
Device
Generic name
Device, Neurovascular Embolization
Manufacturer
Stryker Neurovascular CorkCatalog number
M0036126200
Lot number
25738132
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
DURING COIL EMBOLIZATION, THERE WAS RESISTANCE WHEN ADVANCING THE SUBJECT COIL THROUGH THE SHAFT OF THE MICROCATHETER AND THE SUBJECT COIL DID NOT PROPERLY TRACK THE DELIVERY WIRE DURING REPOSITIONING AND WAS RETRIEVED, AT WHICH POINT COIL STRETCHING WAS OBSERVED. THE REMAINING DISTAL PORTION WAS SUCCESSFULLY RETRIEVED USING A GOOSENECK SNARE WITHOUT IMPACTING THE PROCEDURE. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.