inforMED
MalfunctionQAD

Photodynamic Bone Stabilization system

Received Jun 26, 2026 · Event occurred Jun 1, 2026

Report 3006845464-2026-00002 · MDR key 25635556

Device

Generic name

Intramedullary Fixation Rod, Product Code Qad

Catalog number

SY-2100-01

Product problems

  • Optical Problem

Patient

60 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING AND RESULT WILL BE SUBMITTED UPON COMPLETION.

Description of Event or Problem

TWO ATTEMPTS TO CURE AN ILLUMINOSS IMPLANT RESULTED IN INCOMPLETE POLYMERIZATION WHEN USING THE SAME LED LIGHT BOX, WITH THE FIRST IMPLANT REQUIRING REMOVAL DUE TO FAILURE TO POLYMERIZE. A SECOND IMPLANT ALSO DID NOT FULLY POLYMERIZE AFTER THE INITIAL CURE CYCLE AND REQUIRED EXTENDED CURE TIME (APPROXIMATELY 1800 SECONDS). A REPLACEMENT LED LIGHT BOX WAS SUBSEQUENTLY USED, DEMONSTRATING VISIBLY BRIGHTER LIGHT OUTPUT, AFTER WHICH THE PROCEDURE WAS SUCCESSFULLY COMPLETED.