Received Jun 26, 2026 · Event occurred May 21, 2026

Report 3014312726-2026-00048 · MDR key 25634392

Device

Generic name

Sol-m 10ml Luer Lock Syringe W/o Needle

Model number

P180010

Lot number

UNKNOWN

Product problems

  • Defective Component

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DESPITE MULTIPLE FOLLOW-UPS, THE LOT/BATCH NUMBER COULD NOT BE OBTAINED FROM THE CUSTOMER. CONSEQUENTLY, BATCH RECORD REVIEW AND ARCHIVED SAMPLE ANALYSIS COULD NOT BE PERFORMED. SOL MILLENNIUM RECEIVED THE RETURNED SUBJECT DEVICE FOR EVALUATION. DUE TO CONTAMINATION WITH CYTOSTATIC DRUG RESIDUE, ONLY VISUAL INSPECTION COULD BE PERFORMED. VISUAL INSPECTION CONFIRMED THAT THE ELASTOMERIC STOPPER WAS PARTIALLY DETACHED FROM THE PLUNGER ASSEMBLY. THE RETURNED DEVICE WAS RECEIVED WITH A CLOSED SYSTEM TRANSFER DEVICE (CSTD) ATTACHED. THEREFORE, THE POTENTIAL INFLUENCE OF THE ASSEMBLED SYSTEM, INCLUDING INTERACTION BETWEEN THE SYRINGE AND THE CSTD, CANNOT BE EXCLUDED. AS NO FUNCTIONAL TESTING COULD BE PERFORMED, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. SOL MILLENNIUM WILL CONTINUE TO MONITOR COMPLAINT TRENDS THROUGH ITS POST-MARKET SURVEILLANCE PROCESS.

Description of Event or Problem

CUSTOMER REPORTED THE SYRINGES IN QUESTION, WHICH WERE CONTAMINATED WITH CYTOSTATIC DRUGS. THE FOLLOWING TWO ISSUES HAVE OCCURRED: 1. LIQUID BEHIND THE BLACK STAMP. 2. THE STAMP HAS COME LOOSE FROM THE PLUNGER. IT IS ALSO POSSIBLE THAT THE SCREW-ON ADAPTER FROM SIMPLIVA CAUSED THIS PROBLEM, AS IT MAY NOT HAVE OPENED PROPERLY.