inforMED
MalfunctionFMT

GIRAFFE IRS

Received Jun 26, 2026 · Event occurred May 28, 2026

Report 3008729547-2026-00051 · MDR key 25633406

Device

Generic name

Warmer, Infant Radiant

Catalog number

NA

Lot number

NA

Product problems

  • Mechanical Problem

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN UNAVAILABLE OXYGEN THERAPY. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THE A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 REGULATOR WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.