ENVELLA BED
Received Jun 26, 2026 · Event occurred Jun 2, 2026
Report 1824206-2026-01459 · MDR key 25632772
Device
Generic name
Bed, Air Fluidized
Manufacturer
Baxter Healthcare CorporationModel number
P0819ACatalog number
P0819A
Product problems
- Material Split, Cut or Torn
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE BAXTER TECHNICIAN FOUND THE POWER CORD NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL, IT IS NECESSARY FOR THE ENVELLA AIR FLUIDIZED THERAPY SYSTEM TO HAVE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU DO ANNUAL PREVENTIVE MAINTENANCE (PM). PM CAN HELP MAKE SURE OF A LONG, OPERATIVE LIFE FOR THE UNIT. CHECK THAT THE POWER CORD IS IN GOOD CONDITION: THE PLUG IS A ONE-PIECE MOLDED PLUG ASSEMBLY, THE ASSEMBLY SHOWS NO SIGNS OF CRACKING, THE PLUG MOLDING AROUND THE BLADE IS NOT DISCOLORED, AND THE BLADE IS TIGHT IN THE MOLDING. THE POWER CORD HOOKS SHOW NO SIGN OF CRACKING AND ARE INTACT WITH NO DAMAGE. THE POWER CORD STRAIN RELIEF SHOWS NO SIGN OF CRACKING AND IS INTACT WITH NO DAMAGE. REPAIR OR REPLACE THE PART AS APPLICABLE. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem
A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT ENVELLA BED POWER CORD WAS DAMAGED AND COPPER WIRE EXPOSED. USER DEFINED DESCRIPTION: WAREHOUSE REPAIR. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.