AMS 700 LGX MS PUMP
Received Jun 25, 2026 · Event occurred May 28, 2026
Report 2124215-2026-34056 · MDR key 25630744
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404256Catalog number
72404256
Lot number
1100672748
Product problems
- Fluid/Blood Leak
- Inflation Problem
- Mechanical Problem
- Material Puncture/Hole
- Air/Gas in Device
Patient
57 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100814108. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)(4).
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT FUNCTIONING PROPERLY AND EXHIBITED INFLATION ISSUES. AN ULTRASOUND AND COMPUTED TOMOGRAPHY WERE PERFORMED, AND A FLUID LOSS WAS IDENTIFIED. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE RESERVOIR WAS EXPLANTED AND TESTED, SHOWING NO SOURCE OF LEAK. THE CYLINDERS WERE EXPLANTED AND TESTED SHOWING SMALL LEAK THROUGH SMALL HOLE IN RIGHT CYLINDER, SUSPECTED TO BE RELATED TO A PUNCTURE. ADDITIONALLY, AIRE WAS OBSERVED IN THE DEVICE. BOTH RESERVOIR AND CYLINDERS/PUMP REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.