AMS 700 CX MS PUMP
Received Jun 25, 2026 · Event occurred Jun 16, 2026
Report 2124215-2026-34031 · MDR key 25630697
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404232Catalog number
72404232
Lot number
1100030944
Product problems
- Deflation Problem
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100031370. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)4). THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. DEFLATION ISSUE IS ACKNOWLEDGED WITHIN THE INSTRUCTIONS FOR USE (IFU) AND IS NOT RELATED TO OFF LABEL USE OR FAILURE TO FOLLOW INSTRUCTIONS. BASED ON THE INFORMATION AVAILABLE, THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED; THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.
Description of Event or Problem
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) HAD A DEFLATION ISSUE. A SURGICAL REVISION/REPLACEMENT PROCEDURE WAS PLANNED AS THE NEXT COURSE OF ACTION. THERE WERE NO PATIENT COMPLICATIONS REPORTED.