inforMED
MalfunctionFLE

AMSCO

Received Jun 25, 2026 · Event occurred May 28, 2026

Report 3005899764-2026-00045 · MDR key 25627071

Device

Generic name

Sterilizer

Catalog number

SR0201210311

Product problems

  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT AND FOUND THAT THE VIEWPORT WITHIN THE AMSCO 400 STERILIZER WAS DAMAGED. AS THE VIEWPORT WAS DAMAGED, THIS ALLOWED STEAM TO LEAK OUT DURING A CYCLE. THE TECHNICIAN REPLACED THE VIEWPORT, TESTED THE FUNCTION AND OPERATION OF THE AMSCO 400 STERILIZER, CONFIRMED IT TO BE OPERATING TO SPECIFICATION AND RETURNED THE UNIT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING A CYCLE THEIR AMSCO 400 STERILIZER LEAKED STEAM SUBSEQUENTLY CAUSING THE FACILITY'S FIRE ALARM TO GO OFF. NO REPORT OF INJURY.