WAVELIGHT FS200 FEMTOSECOND LASER
Received Jun 25, 2026 · Event occurred Jun 5, 2026
Report 3003288808-2026-00284 · MDR key 25618058
Device
Generic name
Ophthalmic Laser
Manufacturer
Wavelight GmbhModel number
NACatalog number
8065998284
Lot number
14XWUJ
Product problems
- Unexpected Therapeutic Results
Patient
57 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT THE SURGEON ENCOUNTERED A VERTICAL GAS BREAKTHROUGH AT THE HINGE RESULTED IN INCOMPLETE CUT IN THE RIGHT EYE OF THE PATIENT DURING REFRACTIVE SURGERY. THE SURGEON IMMEDIATELY STOPPED THE PROCEDURE AND RESTARTED USING A NEW PATIENT INTERFACE, PERFORMED A FULL FLAP CUT WITH THE SAME PARAMETERS. THE FLAP WAS SUCCESSFULLY LIFTED, AND THE EXCIMER TREATMENT WAS COMPLETED. NO PATIENT IMPACT HAS BEEN REPORTED AT THIS STAGE. NO FURTHER INFORMATION IS AVAILABLE AT THE TIME.
Additional Manufacturer Narrative
H.3., H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).