inforMED
MalfunctionMEB

FOLFUSOR

Received Jun 24, 2026 · Event occurred May 19, 2026

Report 1416980-2026-02128 · MDR key 25615035

Device

Generic name

Pump, Infusion, Elastomeric

Model number

2C4063K

Catalog number

2C4063K

Lot number

25K019

Product problems

  • Insufficient Flow or Under Infusion

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT A LARGE VOLUME FOLFUSOR HAD A SIGNIFICANT AMOUNT OF FLUID LEFT AFTER AN ADDITIONAL 17HOURS OF THE EXPECTED TIME. THE ISSUE WAS NOTICED AFTER PATIENT USE. THE DEVICE WAS FILLED WITH 4888MG FLUOROURACIL IN 0.9% SODIUM CHLORIDE HAVING A TOTAL VOLUME OF 240ML. THE EXPECTED INFUSION TIME WAS 24HOURS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.