IMMULITE 2000 IGF-I
Received Jun 24, 2026 · Event occurred May 26, 2026
Report 1219913-2026-00124 · MDR key 25607119
Device
Generic name
Radioimmunoassay, Human Growth Hormone
Manufacturer
Siemens Healthcare Diagnostics Products, LimitedCatalog number
11128543
Lot number
357
Product problems
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A CUSTOMER LOCATED OUTSIDE THE UNITED STATES CONTACTED SIEMENS REMOTE SERVICE CENTER REGARDING THE OBSERVATION OF IMMULITE 2000 XPI (SERIAL NUMBER (B)(6)), IMMULITE 2000 IGF-1 (SMN 11128543), LOT 357, GENERATED INITIAL DISCORDANT RESULTS, WHICH WERE REPEATED ON IMMULITE 2000 XPI (SERIAL NUMBER (B)(6)) AND CONSIDERED CORRECT. QUALITY CONTROL MATERIAL WAS IN RANGE. THE ADJUSTMENT IN USE AT THE TIME THE SAMPLES WERE PROCESSED WAS PERFORMED ON (B)(6) 2026 AND DEMONSTRATED ACCEPTABLE PERFORMANCE CONSISTENT WITH THE ANALYZER'S HISTORICAL TREND. THE CUSTOMER RECEIVES SAMPLES FROZEN AT -20°C AND, PRIOR TO ANALYSIS, PERFORMS A THAWING STEP FOLLOWED BY CENTRIFUGATION BEFORE PROCESSING ON THE ANALYZER. AFTER TESTING, SAMPLES ARE REFROZEN TO ALLOW FOR FUTURE REPEAT TESTING IF NECESSARY. DURING THE INVESTIGATION, THE LOCAL SUPPORT TEAM REVIEWED THE ANALYZER EVENT LOGS AND DID NOT IDENTIFY ANY MESSAGES OR ALERTS SUGGESTIVE OF MECHANICAL ISSUES THAT COULD EXPLAIN THE OBSERVED DISCREPANCIES. ADDITIONALLY, THE CUSTOMER DID NOT REPORT SIMILAR OCCURRENCES WITH OTHER ASSAYS PROCESSED ON THE SAME ANALYZER. SIEMENS IS INVESTIGATING.
Description of Event or Problem
THE CUSTOMER REPORTED THE OBSERVATION OF A FALSELY DEPRESSED IGF-I (INSULIN-LIKE GROWTH FACTOR I) RESULT OBTAINED ON ONE (1) PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT. THE SAMPLE WAS REPEATED ON AN ALTERNATE IMMULITE 2000 XPI INSTRUMENT. THE REPEATED RESULT WAS CONSIDERED AS CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED IGF-I RESULT.