COBAS B 221 <6> SYSTEM

Received Jun 24, 2026 · Event occurred Jun 7, 2026

Report 1823260-2026-02509 · MDR key 25606758

Device

Generic name

Blood Gas Analyzer

Manufacturer

Roche Diagnostics

Catalog number

03337154001

Product problems

  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ELECTRODE LOT NUMBERS AND EXPIRATION DATES WERE REQUESTED, BUT NOT PROVIDED. MEDWATCH FIELD D2B PROCODE: PRODUCT CODE KHP APPLIES TO THE THB PARAMETER. OTHER PRODUCT CODES ASSOCIATED WITH THE COBAS B 221 <6> SYSTEM ARE CEM, CDS, CGA, GGF, GGZ, GHS, GKR, JFP, JGS, KHG, AND KHP. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED FOR THB, SO2, AND HCT ON TWO COBAS B 221 <6> SYSTEM ANALYZERS. THE SAMPLE INITIALLY RESULTED IN THE FOLLOWING VALUES WHEN TESTED ON COBAS B 221 SERIAL NUMBER: (B)(6): THB = 6.20 G/DL WITH A DATA FLAG, 10.51 G/DL, AND 13.05 G/DL SO2 = 97.1%. HCT = 21.9 % WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON COBAS B 221 SERIAL NUMBER: (B)(6), RESULTING IN THE FOLLOWING VALUES: THB = 8.73 G/DL WITH A DATA FLAG. SO2 = 90.1% WITH A DATA FLAG. HCT = 33.8 % WITH A DATA FLAG. THE THB VALUE OF 6.20 G/DL WAS CONSIDERED CORRECT.