inforMED
MalfunctionLGD

ELECSYS TOXO IGG

Received Jun 24, 2026 · Event occurred May 22, 2026

Report 1823260-2026-02501 · MDR key 25606608

Device

Generic name

Enzyme Linked Immunoabsorbent Assay, Toxoplasma Gondii

Manufacturer

Roche Diagnostics

Catalog number

04618815190

Lot number

86681601

Product problems

  • High Test Results
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 3 PATIENTS' SAMPLES TESTED WITH ELECSYS TOXO IGG ASSAY ON A COBAS E 411 ANALYZER (DISK SYSTEM). PATIENT 1: INITIAL RESULT: 0.926 IU/ML. REPEAT RESULT: 650 IU/ML (ACCOMPANIED BY A DATA FLAG). PATIENT 2: INITIAL RESULT: 0.452 IU/ML. REPEAT RESULT: 176.02 IU/ML. PATIENT 3: INITIAL RESULT: 0.800 IU/ML. REPEAT RESULT: 650 IU/ML (ACCOMPANIED BY A DATA FLAG).

Additional Manufacturer Narrative

THE COBAS E 411 ANALYZER (DISK SYSTEM) SERIAL NUMBER WAS (B)(6). THE QC WAS ACCEPTABLE. THE INVESTIGATION IS ONGOING.