AMBU AURASTRAIGHT U SIZE 4

Received Jun 24, 2026

Report 9610691-2026-00021 · MDR key 25603752

Device

Generic name

Ambu Aurastraight U Size 4

Manufacturer

Ambu A/s

Catalog number

324400000U

Product problems

  • Improper or Incorrect Procedure or Method

Patient

NA · Unknown

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

Narrative

Additional Manufacturer Narrative

THIS COMPLAINT HAS BEEN RE-EVALUATED AS PART OF A CORRECTIVE ACTION RELATED TO AN OBSERVATION ON RISK MANAGEMENT IDENTIFIED DURING AN FDA INSPECTION. H6: HEATH EFFECT CODE: NO CODE AVAILABLE FOR "CUTTING PATIENT'S THROAT". INVESTIGATION: NEITHER SAMPLE NOR PICTURE FOR INVESTIGATION, VISUAL INSPECTION IS NOT PERFORMED, THE CUSTOMER REPORTED FAILURE CANNOT BE VERIFIED. PER THE CUSTOMER REPORTED THAT THE PRODUCT CUTTING PATIENT'S THROATS, IT MAY BE CAUSED BY THE EXCESSIVE FORCE DURING INSERTION OR MAY BE CAUSED BY SHARP EDGES ON THE PRODUCT. DUE TO LIMITED INFORMATION, THE ROOT CAUSE CANNOT BE DEFINED. THE PATIENT WAS INVOLVED, BUT THE PATIENT CONDITION IS UNKNOWN, THE COMPLAINT RATE IS LOW. DUE TO THE ROOT CAUSE THAT CANNOT BE DEFINED AS LIMITED INFORMATION, THE CORRECT ACTION WILL NOT BE TAKEN AT THIS TIME. SINCE THE ROOT CAUSE CANNOT BE IDENTIFIED, THERE IS NO ADDITIONAL ACTION IS NOT REQUIRED AT THIS MOMENT.

Description of Event or Problem

WE HAVE BEEN RECEIVING A LOT OF PROVIDER COMPLAINTS ABOUT THE LMA'S CUTTING PATIENT'S THROATS. SPECIFICALLY THE 3244.