ALINITY S CHAGAS REAGENT KIT

Received Jun 23, 2026 · Event occurred Apr 5, 2026

Report 1415939-2026-00016 · MDR key 25601952

Device

Generic name

Serology Tests Used For Infectious Disease Screening Of Blood And Plasma Donatio

Catalog number

06P08-61

Lot number

75047MM00

Product problems

  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.

Description of Event or Problem

THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY S CHAGAS AND PROVIDED THE FOLLOWING DATA: SAMPLE ID: (B)(6) ¿ FEMALE, 19 ON (B)(6), 2026, THE RESULTS WERE 1.02 S/CO, 1.02 S/CO, AND 1.06 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS NEGATIVE. SAMPLE ID: (B)(6) ¿ MALE, 17 ON (B)(6) 2026, THE RESULTS WERE 2.88 S/CO, 2.86 S/CO, AND 2.86 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6) ¿ MALE, 35 ON (B)(6), 2026, THE RESULTS WERE 1.22 S/CO, 1.18 S/CO, AND 1.17 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS NEGATIVE. SAMPLE ID: (B)(6)¿ MALE, 19 ON (B)(6), 2026, THE RESULTS WERE 1.73 S/CO, 1.76 S/CO, AND 1.68 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6)¿ MALE, 39 ON (B)(6), 2026, THE RESULTS WERE 1.87 S/CO, 1.85 S/CO, AND 1.85 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6) ¿ MALE, 19 ON (B)(6), 2026, THE RESULTS WERE 2.16 S/CO, 2.19 S/CO, AND 2.10 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6)¿ MALE, 56 ON (B)(6), 2026, THE RESULTS WERE 1.27 S/CO, 1.30 S/CO, AND 1.30 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS NEGATIVE. SAMPLE ID: (B)(6)¿ MALE, 17 ON (B)(6), 2026, THE RESULTS WERE 1.55 S/CO, 1.53 S/CO, AND 1.48 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6) ¿ FEMALE, 38 ON (B)(6), 2026, THE RESULTS WERE 1.01 S/CO AND 1.04 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS INDETERMINATE. SAMPLE ID: (B)(6). ON (B)(6), 2026, THE RESULTS WERE 9.77 S/CO, 10.19 S/CO, AND 9.27 S/CO WITH LOT 75047MM00 ON ANALYZER (B)(6); CONFIRMATORY TESTING WAS NOT AVAILABLE. THE CONFIRMATORY TEST UTILIZED FOR CHAGAS WAS CESA. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.