NEUROFORM ATLAS 4.5MM X 15MM NO TIP - CE
Received Jun 22, 2026 · Event occurred May 28, 2026
Report 3008881809-2026-00381 · MDR key 25587138
Device
Generic name
Intracranial Coil-assist Stent
Manufacturer
Stryker Neurovascular CorkCatalog number
M003EZAS45150
Lot number
25923330
Product problems
- Premature Activation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING PCOMA ANEURYSM COIL EMBOLIZATION PROCEDURE, WHEN THE SUBJECT STENT WAS ADVANCING THROUGH THE MICROCATHETER, THE SUBJECT STENT GOT STUCK IN THE MIDDLE OF THE MICROCATHETER. THE PHYSICIAN WITHDREW THE SUBJECT STENT TOGETHER WITH THE MICROCATHETER OUT OF THE BODY, SWITCHED TO A DIFFERENT BRAND OF MICROCATHETER AND STENT, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, DURING INSPECTION ON THE TABLE, WHEN RETRACTING THE DELIVERY WIRE, IT WAS FOUND THAT THE SUBJECT STENT HAD DEPLOYED WITHIN THE MICROCATHETER. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.