inforMED
MalfunctionDRG

ORG-9110A

Received Jun 22, 2026 · Event occurred Jun 1, 2026

Report 8030229-2026-06686 · MDR key 25586002

Device

Generic name

Multiple Patient Receiver

Model number

ORG-9110A

Catalog number

ORG-9110A

Lot number

NA

Product problems

  • Communication or Transmission Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE THIS MULTIPLE PATIENT RECEIVER (ORG) GOING INTO COMM LOSS FOR ABOUT 30 SECONDS AND THEN COMES BACK. CUSTOMER STATES THAT THIS SEEMS RANDOM, APPROXIMATELY EVERY 30MIN OR SO. THIS IS AFFECTING ONLY ONE ORG, BED-ID'S TELE20- TELE26. ALL 7 BEDS ARE GOING INTO COMM LOSS AT THE SAME TIME (THIS ORG IS ONLY SETUP FOR 7 BEDS, THERE IS A FREE, UNUSED CHANNEL THAT IS NOT BEING MONITORED). NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAVE THIS MULTIPLE PATIENT RECEIVER (ORG) GOING INTO COMM LOSS FOR ABOUT 30 SECONDS AND THEN COMES BACK. CUSTOMER STATES THAT THIS SEEMS RANDOM, APPROXIMATELY EVERY 30MIN OR SO. THIS IS AFFECTING ONLY ONE ORG, BED-ID'S TELE20- TELE26. ALL 7 BEDS ARE GOING INTO COMM LOSS AT THE SAME TIME (THIS ORG IS ONLY SETUP FOR 7 BEDS, THERE IS A FREE, UNUSED CHANNEL THAT IS NOT BEING MONITORED). NO PATIENT HARM WAS REPORTED.