inforMED
MalfunctionJFT

ELECSYS CORTISOL II

Received Jun 22, 2026 · Event occurred May 27, 2026

Report 1823260-2026-02458 · MDR key 25575734

Device

Generic name

Fluorometric, Cortisol

Manufacturer

Roche Diagnostics

Catalog number

06687733190

Lot number

922327

Product problems

  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE EXPIRATION DATE FOR THE ELECSYS CORTISOL II REAGENT WAS NOT PROVIDED. THE COBAS E 411 ANALYZER (DISK SYSTEM) SERIAL NUMBER WAS (B)(6). THE PATIENT SAMPLE IS NOT AVAILABLE FOR INVESTIGATION. THE CALIBRATION WAS PERFORMED ON (B)(6) 2026, AND THEY WERE ACCEPTABLE. THE QC WAS WITHIN THE RANGES. THE ALARM TRACE SHOWED 2 "LIQUID FLOW CLEANING (LFC) IS RECOMMENDED" ALARMS AND 2 SAMPLE PIPETTING ALARMS ON (B)(6) 2026. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS CORTISOL II ASSAY ON A COBAS E 411 ANALYZER (DISK SYSTEM). THE INITIAL RESULT WAS >1750 NMOL/L (ACCOMPANIED BY A DATA FLAG). THE 1ST REPEAT RESULT WAS >1750 NMOL/L. THE SAMPLE WAS THEN SENT TO A DIFFERENT LABORATORY, AND ON (B)(6) 2026, THE 2ND REPEAT RESULT WAS 291 NMOL/L. THE REPEAT RESULT OF 291 NMOL/L WAS CONSIDERED "NORMAL". AN INTERFERENT IS SUSPECTED IN THE PATIENT SAMPLE.