EVIS LUCERA ELITE XENON LIGHT SOURCE
Received Jun 19, 2026 · Event occurred Mar 20, 2026
Report 3002808148-2026-21091 · MDR key 25564401
Device
Product problems
- Corroded
- Gas/Air Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE COMPLAINT WAS TRACED TO COMPONENT FAILURE SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
THE CUSTOMER RETURNED THE XENON LIGHT SOURCE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPORT TECHNICIAN INDICATES THAT INTERNAL LIQUID IMMERSION AND RUST IN FRONT PANNEL WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.