MOBILETT MIRA MAX
Received Jun 19, 2026 · Event occurred Jun 3, 2026
Report 3004977335-2026-00108 · MDR key 25559659
Device
Product problems
- Component Missing
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
E1: THE REPORTER¿S CONTACT NAME AND PHONE NUMBER WERE NOT PROVIDED FOR REPORTING PURPOSES. H3, H6: INITIAL ACTIONS (CORRECTIVE AND/OR PREVENTIVE): NO GENERAL PROBLEM HAS BEEN DETECTED FOR THE INSTALLED BASE WHICH REQUIRES IMMEDIATE ACTION. MANUFACTURER'S PRELIMINARY RESULTS AND CONCLUSION: THE ROOT CAUSE IS CURRENTLY UNKNOWN AND IS UNDER INVESTIGATION. SIEMENS HEALTHINEERS WILL SUBMIT A SUPPLEMENTAL REPORT TO THE FDA UPON COMPLETION OF THE INVESTIGATION.
Description of Event or Problem
SIEMENS HEALTHINEERS WAS NOTIFIED OF AN ISSUE WITH THE MOBILETT MIRA MAX SYSTEM. IT WAS COMMUNICATED THAT THE GROUNDING CONNECTION IN THE CABLE WINCH PLUG IS INSUFFICIENT. USUALLY, THIS KIND OF PLUG (FRENCH-TYPE POWER PLUG) IS EQUIPPED WITH TWO METAL TABS WHICH CONTACT WITH THE GROUNDING PRONGS ON THE SOCKET. THIS ALLOWS THE ENTIRE METAL STRUCTURE TO CONNECT TO THE GROUNDING WIRE. IN THIS CASE ONLY ONE OF THESE METAL TABS WAS AVAILABLE AND SO THE GROUNDING FUNCTION WAS NOT FULLY ENSURED. IN CASE OF A SECOND FAULT CONDITION AT THE SYSTEM, THE CURRENT LEAKAGE CANNOT BE PROPERLY DISCHARGED THROUGH THE GROUND WIRE AND THUS AN ELECTRIC SHOCK MIGHT BE POSSIBLE. THERE WAS NO PATIENT OR STAFF INJURY ASSOCIATED WITH THIS ISSUE. ALTHOUGH THERE WAS NO INJURY ASSOCIATED WITH THE COMPLAINT ISSUE, IN A WORST-CASE SCENARIO, A SERIOUS INJURY OR DEATH COULD RESULT DUE TO ELECTRIC SHOCK.