inforMED
MalfunctionOJA

SUGAN CO., LTD..OSAKA..JP.ETO.CUSTOM ANGIOGRAPHIC KIT.

Received Jun 18, 2026 · Event occurred Mar 24, 2026

Report 1721504-2026-00256 · MDR key 25551930

Device

Generic name

Intravenous Extension Tubing Set

Catalog number

K08-MP5208

Lot number

H3419378

Product problems

  • Contamination

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED UNIT FOUND THE PACKAGING TO BE UNOPENED AND IN GOOD CONDITION. UPON EXAMINING THE BLUE STRIPED TUBING IN QUESTION, A FOREIGN OBJECT WAS ABLE TO BE CONFIRMED LOOSE WITHIN THE FLUID PATHWAY. THE COMPLAINT WAS CONFIRMED. THE ROOT CAUSE IS ATTRIBUTABLE TO THE ENVIRONMENT IN WHICH THE PRODUCT IS MANUFACTURED. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE PRODUCT HISTORY WAS REVIEWED, AND NO DEVIATIONS, NON-CONFORMANCES, OR EXCEPTION DOCUMENTS WERE FOUND. NONCONFORMANCES OF THIS NATURE ARE MONITORED BY THE OPERATION TEAM. THIS COMPLAINT WILL BE USED TO TREND FOR SIMILAR COMPLAINTS.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A RED FOREIGN OBJECT IN THE BLUE STRIPED TUBE.